ZHAO Pei-tao, ZHOU Jian-kun, CAO Kai-fa, HU Shun-xing. Calibration Experiment of Vibrational Raman Lidar for Bottom Troposphere Carbon Dioxide Profile DetectingJ. Transactions of Beijing institute of Technology, 2015, 35(6): 616-620. DOI: 10.15918/j.tbit1001-0645.2015.06.014
Citation: ZHAO Pei-tao, ZHOU Jian-kun, CAO Kai-fa, HU Shun-xing. Calibration Experiment of Vibrational Raman Lidar for Bottom Troposphere Carbon Dioxide Profile DetectingJ. Transactions of Beijing institute of Technology, 2015, 35(6): 616-620. DOI: 10.15918/j.tbit1001-0645.2015.06.014

Calibration Experiment of Vibrational Raman Lidar for Bottom Troposphere Carbon Dioxide Profile Detecting

  • Based on the principle of Raman scatting, the lower troposphere CO2 profile monitoring vibrational Raman lidar system was developed. Combined with the use of three-dimensional optical scanning system, the lidar has the three-dimensional scanning function, which can realize the spatial distribution of atmospheric carbon dioxide profile monitoring. Combined with the lidar characteristic, a new calibration method of vibrational Raman lidar was presented, and the lidar system calibration experiments were carried out to obtain the profile line of carbon dioxide Hefei area. The calibration uncertainty of the experiments was analyzed and obtained the uncertain degree is 1%, the result proves the reliability of the calibration experiment method.
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return